Technical Specifications - Medtronic GUARDIAN 4 Manual Del Usuario

Ocultar thumbs Ver también para GUARDIAN 4:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 77
Disposal
Do not dispose of the transmitter, charger, and tester in unsorted municipal waste stream. Dispose of
the transmitter, charger, and tester according to local regulations for electronic waste disposal.

Technical Specifications

Table 2. Product specifications
Biocompatibility
Applied parts
Operating conditions
Storage conditions
Battery life
Transmitter frequency 2.4 GHz band, Bluetooth
Effective radiated
power (ERP)
Effective isotropic
radiated power (EIRP)
Operating range
Transmitter expected
service life
Transmitter: Complies with EN ISO 10993-1
Transmitter
Sensor
Transmitter temperature: 0°C to 45°C (32°F to 113°F)
Caution: When operating the transmitter on a tester in air temperatures
greater than 41°C (106°F), the temperature of the transmitter may exceed
43°C (109°F).
Transmitter relative humidity: 10% to 95% with no condensation
Transmitter pressure: 57.60 kPa to 106.17 kPa (8.4 psi to 15.4 psi)
Charger temperature: 10°C to 40°C (50°F to 104°F)
Charger relative humidity: 30% to 75% with no condensation
Transmitter temperature: -20°C to 55°C (-4°F to 131°F)
Transmitter relative humidity: up to 95% with no condensation
Transmitter pressure: 57.6 kPa to 106 kPa (8.4 psi to 15.4 psi)
Charger temperature: -10°C to 50°C (14°F to 122°F)
Charger relative humidity: 10% to 95% with no condensation
Transmitter: Seven days of continuous glucose monitoring immediately
following a full charge.
Charger: The charger uses one new AAA battery to charge the transmitter.
-12.05 dBm (0.06 mW)
-9.9 dBm (0.1 mW)
Up to 1.8 meters (6 feet) in free-air
The transmitter expected service life is one year depending on patient
usage.
37