• Confirm you are matching the correct range. Control Solution 1 results
should be matched to the CTRL 1 range printed on the test strip vial (or
on the foil pouch).
• Check the expiration date of the test strip and control solution. Make
sure that the test strip vial and control solution bottle have not been
opened for more than 6 months. Discard any test strips or control
solution that has expired.
• Confirm the temperature in which you are testing is between 10 and
40 °C (50-104°F).
• Make sure that the test strip vial and control solution bottle have been
tightly capped.
• Confirm that you are using the same brand of control solution that was
provided with your kit.
• Make sure that you followed the test procedure correctly.
After checking all of the conditions listed above, repeat the quality control test
with a new test strip. If your results still fall outside of the control range shown
on the test strip vial (or on the foil pouch), your meter may be defective.
Contact your local distributor for help.
Three
levels of control solution are available labeled
Control Solution 1 and Control Solution 2. Control Solution 1 is sufficient for
most all self-testing needs. If you think your meter or strips may not be working
correctly, you may also want to do
0, CTRL 1 and CTRL 2
pouch). Simply repeat step 4 through 6, using
Solution 2.
For confirmation of results, C
CTRL 0 range,
Control Solution 1 tests should fall within the CTRL 1 range,
and Control Solution 2 tests should fall within the CTRL 2 range. If the control
solution test results do not fall within the respective ranges, DO NOT use the
system to test blood, as the system may not be working properly. If you cannot
fix the problem, contact your local distributor for help.
Please contact your local distributor for information on ordering the On Call
Vivid Glucose Control Solution kit, which contains
Solution 1 and Control Solution 2.
All manuals and user guides at all-guides.com
level 0 or
are displayed on the test strip vial (or on the foil
ontrol Solution 0 tests should fall within the
19
Control Solution 0,
level 2 test. The ranges for
Control Solution 0 or
Control Solution 0,
CTRL