Display
Temperature
operating temperature of the system.
Temperature is below the operating
temperature of the system
Battery is low but has enough power
to run 20 more tests.
Battery is discharged and meter does
not
allow
for
replacement with a new battery.
Meter electronics failure.
Improper type of strip used.
Communications failure.
Strip testing error.
Meter data port is plugged in with
data transfer cable when meter is in
waiting for sample application mode
with strip already inserted into the
meter's strip port.
All manuals and user guides at all-guides.com
Causes
has
exceeded
the
Move to a cooler environment and repeat the test.
Move to a warmer environment and repeat the test.
Test results will still be accurate, but replace the
battery as soon as possible.
more
tests
until
Replace the battery and repeat the test
If the problem persists, contact your local distributor.
Damaged test strip or calibration error. Please test
again by using a new strip. If problem persist,
please contact your local distributor.
There is an error in transferring data to the PC. See
the package insert included with the On Call
Diabetes
troubleshooting.
Repeat test and apply enough sample to fill the test
strip check window within 3 seconds. When repeat
testing, do not touch the strip during meter count
down. Please make sure fresh blood sample with
intended hematocrit level is used. Please make sure
blood sample is not contaminated. If problem
persists, please contact your local distributor.
Unplug the data transfer cable from meter data port,
and then remove the strip and insert the strip to the
strip port for testing. If the problem persists, contact
your local distributor.