Troubleshooting Guide

The meter has built-in messages to alert you of problems. When error
messages appear, note the error number, turn off the meter and then follow
these instructions.
Display
Battery may be damaged or empty
Meter fails
to turn on
Meter is too cold
Power On self check error
Internal calibration check error
Test strip was removed during the
test.
Sample was applied to the test strip
too soon.
Test strip is contaminated or used.
Sample was applied to the test strip
too soon.
Insufficient sample.
Sample application error due to late
sample re-dosing.
All manuals and user guides at all-guides.com
Causes
Replace battery.
If meter has been exposed to or stored in cold
conditions, wait 30 minutes to allow meter to reach
room temperature then repeat test.
Remove battery for 30 seconds and then put battery
back and turn meter on again. If problem persists,
contact your local distributor.
If a cell phone, radio frequency source or a high
power electrical source is nearby, place more
distance between the meter and any of these
sources then retest. If the problem persists, contact
your local distributor.
Repeat the test and ensure test strip remains in
place.
Repeat test and apply sample after blood drop/test
strip symbol appears.
Repeat test with a new test strip.
Repeat test and apply sample after blood drop/test
strip symbol appears.
Repeat test and apply enough sample to fill the test
strip check window.
Repeat test and apply enough sample to fill the test
strip check window within 3 seconds.
41
Solution