7.11 Animal weighing function

The balance has an integrated animal weighing function (mean value calculation).
With this function it is possible to weigh domestic or small animals exactly, although
they do not stand quiet on the weighing plate.
Note: If they move too much, an exact weighing will not be possible.
In the menu the animal weighing function can be switched on or off. To achieve this,
follow the sequence of operations below:
Switch-on the balance using the
Press the
key and keep it pressed until „UNIT" appears on the display.
Press the
key eight times, in the display appears „ANL".
Confirm by pressing the
Use the
key to choose one of the following settings:
Display
„ANL" off
„ANL" 3
„ANL" 5
„ANL" 10
„ANL" 15
Use the
key to confirm the selected setting.
Operation:
Switch-on the balance using the ON key and wait for the „0" display.
Put the weighing good (animal) on the weighing plate and press the
display the preselected time is displayed in seconds and then is counted towards
zero. During this time the balance takes up several measuring values. When
reaching „0" sounds an acoustic signal and the calculated weighing value is
displayed.
By pressing the
mode.
Repeated pressing of the
22
key and wait for the „0" display.
key.
Function
Animal weighing function is switched off
Weighing value calculation above 3 sec. till to the value
display
Weighing value calculation above 5 sec. till to the value
display
Weighing value calculation above 10 sec. till to the value
display
Weighing value calculation above 15 sec. till to the value
display
key several times, the balance returns to the normal weighing
key activates this function anew.
key. In the