► Take suitable measures if required (e.g. cleaning, repair,
replacement, inspection by the manufacturer or a specialist
workshop etc.).
CAUTION
Exceeding the service life
Risk of injury due to change in or loss of functionality and damage to
the product
► Ensure that the approved service life is not exceeded.
CAUTION
Mechanical damage to the product
Risk of injury due to change in or loss of functionality
► Use caution when working with the product.
► If the product is damaged, check it for proper function and readi­
ness for use.
► In case of changes in or loss of functionality, do not continue
using the product (see "Signs of changes in or loss of functional­
ity during use" in this section).
► Take any necessary measures (e.g. repair, replacement, inspec­
tion by the manufacturer's customer service, etc.).
Signs of changes in or loss of functionality during use
Among other factors, changes in functionality can be indicated by an
altered gait pattern, a change in the positioning of the prosthetic com­
ponents relative to each other and by the development of noises.
4 Scope of delivery
Quantity
Designation
1

Instructions for Use

1
Sliding Tube Clamp
Adapter
1
Cap Screw
8
Reference number
501Z2=M6x25
Quantity
Designation
1
Cap Screw
4
Set Screw
5 Preparing the product for use
CAUTION
Incorrect alignment or assembly
Risk of injury due to damaged prosthetic components
► Observe the alignment and assembly instructions.
CAUTION
Improper assembly of the screw connections
Risk of injury due to breakage or loosening of the screw connec­
tions
► Clean the threads before every installation.
► Apply the specified torque values.
► Follow the instructions regarding the length of the screws and
about how to secure the screws.
CAUTION
Incorrect mounting of the tube
Risk of injury due to breakage of load-bearing components
► Slide the tube all the way to the stop in the intended prosthetic
component when mounting.
5.1 Mounting the adapter
Required materials: torque wrench (e.g. 710D20), degreasing
>
cleaner (e.g. acetone), 636K13 Loctite 241
1) Loosen the tube clamp cap screw by two turns.
2) For final assembly: Degrease the insertion range of the tube
clamp adapter and tube adapter.
Reference number
501Z2=M6x35
506G3=M8x16