Horizontal mounting
Abb. 10: Horizontal mounting
4
Connecting to power
supply
4.1

Preparing the connection

Note safety instructions
Always keep in mind the following safety
instructions:
Connect only in the complete absence of
line voltage
Take note of safety instructions for Ex
applications
Connect the voltage supply according to the
following connection diagrams. The electronics
module with relay output is designed in protec-
tion class I. To maintain this protection class, it is
absolutely necessary that the ground conductor
be connected to the internal ground terminal.
The data for power supply are specified in chap-
ter "Technical data".
Connection cable
VEGAVIB S 61 is connected with standard
cable with round cross section. An outer cable
diameter of 5 ... 9 mm (0.2 ... 0.35 in) ensures
the seal effect of the cable gland.
Make sure that the cable used has the required
temperature resistance and fire safety for max.
occurring ambient temperature
If cable with a different diameter or wire cross
section is used, exchange the seal or use an
appropriate cable connection.
4.2

Connection procedure

Proceed as follows:
VEGAVIB S 61 •
20°

4 Connecting to power supply

1. Unscrew the housing lid
2. Loosen compression nut of the cable gland
and remove blind plug
3. Remove approx. 10 cm (4 in) of the cable
mantle, strip approx. 1 cm (0.4 in) of insula-
tion from the ends of the individual wires
4. Insert the cable into the sensor through the
cable entry
5. Lift the opening levers of the terminals with
a screwdriver (see following illustration)
6. Insert the wire ends into the open terminals
according to the wiring plan
7. Press down the opening levers of the
terminals, you will hear the terminal spring
closing
8. Check the hold of the wires in the terminals
by lightly pulling on them
9. Tighten the compression nut of the cable
entry gland. The seal ring must completely
encircle the cable
10. If necessary, carry out a fresh adjustment
11. Screw the housing lid back on
The electrical connection is finished.
Abb. 11: Connection steps 5 and 6
17