Descargar Imprimir esta página

Agilent Technologies TwisTorr 74 FS Serie Manual De Instrucciones página 310

Publicidad

17
Technical Information
Inlet screen installation
prevent the rotor blades from being damaged by debris. The inlet
screen, however, does reduce the pumping speed as per following
table:
X3502-68000
X3502-68001
The inlet screen is fitted in the upper part of the pump, as shown in
the figure.
Figure 30
310/342
Reduction in pumping speed (%)
Ar
N
2
20%
18%
TwisTorr 74 FS User Manual / 87-901-053-01
He
H
2
7%
5%

Publicidad

loading

Este manual también es adecuado para:

X3502-64170X3502-64171X3502-64172X3502-64173