17
Technical Information
Inlet screen installation
prevent the rotor blades from being damaged by debris. The inlet
screen, however, does reduce the pumping speed as per following
table:
X3502-68000
X3502-68001
The inlet screen is fitted in the upper part of the pump, as shown in
the figure.
Figure 30
310/342
Reduction in pumping speed (%)
Ar
N
2
20%
18%
TwisTorr 74 FS User Manual / 87-901-053-01
He
H
2
7%
5%