Currents (A Dc And Ac) - Chauvin Arnoux C.A 5003 Manual De Empleo

Ocultar thumbs Ver también para C.A 5003:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

5 - DC AND AC CURRENTS
Always switch off the circuit to test before connecting the
multimeter to the circuit. If the "Fus." light comes on, change the
faulty fuse(s) (Reminder: minimum voltage of 100 V).
Connect the leads to the multimeter and connect in series to the
circuit with:
- the red lead in the "mmA" terminal, up to 1.5 mA
- the red lead in the "15 A" terminal, from 1.5 A to 15 A
Switch off the power supply to the circuit before changing range.
When the order of magnitude is not known, place the switch on the
highest range then progressively lower to the appropriate range.
To get the current in mA, mA or A, multiply the value read on the
appropriate scale by the reading coefficient shown in the table.
5 0 µ A (1) 1 5 0 µ A
A D C
S c a le
R e a d in g
c o e ffic ie n t
V o lta g e d ro p
a t th e
te rm in a ls (2)
A c c u ra c y (3)
P ro te c tio n
Limitation 10 min On, 5 min Off up to +40° C max
(1) Common to the 100 mV DC range
(2) Without the leads. Resistance of the pair of leads supplied: approx. 70 mW
(3) In % of the end of scale
Do not use the A AC input on unprotected current transformers.
A A C (1)
S c a le
R e a d in g
c o e ffic ie n t
V o lta g e d ro p
a t th e
te rm in a ls (2)
A c c u ra c y (3)
P ro te c tio n
Limitation 10 min On, 5 min Off up to +40° C max
(1) Bandwidth: 40 Hz to 5 kHz
(2) Without the leads. Resistance of the pair of leads supplied: approx. 70 mW
1 4
1 0 0
x 0 .5
x 1 0
< 0 .1 V
1 .5 m A
1 5 m A
x 0 .1
< 0 .8 V
1 .5 m A
1 5 m A
1 5 b la c k
x 0 .1
x 1
< 0 .2 5 V
2 .5 %
F u s e 1 .6 A H B C
1 5 0 m A
1 5 re d
x 1
x 1 0
2 .5 %
F u s e 1 .6 A H B C
(3) In % of the end of scale
1 5 0 m A
1 .5 A
x 1 0
x 0 .1
< 0 .3 V
< 0 .8 V
1 .5 A
1 5 A
x 0 .1
< 0 .9 V
1 5 A
x 1
5 %
Fuse 16A
H BC
x 1
< 0 .8 V
5 %
Fuse 16 A
H B C

Publicidad

Tabla de contenido
loading

Tabla de contenido