The device is intended for use in. the electromagnetic environment specified below. The customer or
the user should assure that it is used in such an environment.
Immunity test
Immunity test
Conducted RF
3 Vrms 150 kHz to 80
MHz
lEC 61000-4-6
Radiated RF lEC
3 V/m 80 MHz to 2.5
GHz
61000-4-3
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, (a)should
be less than the compliance level in each frequency range. (b)Interference may occur In the vicinity of
equipment marked with the following symbol
Recommended separation distances between portable and mobile RF communications
IACER S r
Electromagnetic immunity a r.f.
Immunity test
3 V
3 V/m
53 de 56
typical commercial or
hospital environment.
Immunity test
Portable and mobile RF
Communications
equipment should be
used no closer to any
part of the device,
including cables, than
the recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
separation distance
d= 1.2 √P
d= 1.2 √P for 80 MHz
to 800 MHz
d= 2.3√P for 800 MHz
to 2.5 GHz
Where P is the
maximum output power
rating of the transmitter
In watts (W) according
to the. Transmitter
manufacturer and d is
the recommended
separation distance in
meters (m).
PG229 Rev 00 17 06 16