Electromagnetic immunity
The device is inteded for use in the electromagnetic environment specified below. The customer or the
user of the device should assure that is used in suche environment
Immunity test
Electrostatic
discharge
(ESD)
EN 61000-4-2
Mains
power
electromagnetic field
EN 61000-4-8
Guidance and manufacturer's declaration – electromagnetic immunity
The device is inteded for use in the electromagnetic environment specified below. The customer or
the user of the device should assure that is used in suche environment
Immunity test
Test level EN
60601-1-2
Conducted RF
3 Vrms 150kHz
EN 61000-4-6
to 80MHz
RF Radiata
3 Vrms 80MHz
EN 61000-4-3
to 2,5GHz
Field strangths from fixed RF transmitters, are determined by an electromagnetic site survey, should
be less than the complicance level in each frequency rage.
Interference may occur in the vicinity of equipment marked with the following symbol:
Recommended separation distances between portable and mobile communications
equipment and the device
The device is intended for the use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the device can help prevent electromagnetic
interferences by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the device as recommended below, according to the maximum output
power of the communication equipment.
I.A.C.E.R. Srl
Test level EN
Compliance level
60601-1-2
 6kV contact
 6kV contact
 8kV air
 8kV air
3 A/m
3 A/m
Compliance level
3 Vrms 150kHz to
80MHz
3 Vrms 80MHz to
2,5GHz
22 de 24
used for domestic purposes.
Electromagnetic environment
– guidance
Floors sholud be wood, concrete
or ceramic tile. If floor are
covered with syntethic material,
the relative humidity should be
at least 30%
Mains power quality should be at
that of a typical commercial or
hospital environment
Electromagnetic
environment
guidance
Portable
and
communications equipment should be
used no closer to any part of the device,
including
cables,
recommended
separation
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance:
d = 1,2 P 150kHz to 80MHz
d = 1,2 P 80 MHz to 800 MHz
d = 2,3 P 800 MHz to 2,5 GHz
where (P) is the maximum output
power rating of the transmitter in Watts
(W) according to the transmitter
manufacturer
and
recommended separation distance in
metres (m).
MNPG157-05 de 03/12/2015