Cleaning · Autoclaving
When properly used, the instrument is maintenance-free. Before auto-
claving, the instrument must be cleaned carefully. Wipe the housing with
a damp cloth. The instrument is autoclavable at 121 °C (250 °F), 2 bar
absolute (30 psi) with a holding time of at least 15 minutes according to
DIN EN 285.

Replacing the Filter and Cleaning the Adapter

Only necessary if a pipette has been overfilled or
the suction capacity is impaired.
Warning!
Wear protective gloves and eye protection.
Avoid the risk of contamination!
1.
Pull out the pipette.
2.
Unscrew the adapter support.
3.
Pull out the membrane filter.
4.
Remove the adapter from the adapter
support by pushing it upwards.
5.
To clean adapter, carefully rinse it by
using an appropriate compatible solvent
(e.g., water) in a wash bottle.
Blow out the liquid and allow to dry
completely.