Duty Cycle - Chauvin Arnoux C.A 5233 Manual De Instrucciones

Ocultar thumbs Ver también para C.A 5233:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

English
4.2.3
Alternating voltages at low impedance (V
AC LowZ True RMS)
The input impedance is 270kΩ.
A low input impedance serves to eliminate the effects of
interference voltages due to the supply network, and makes it
possible to measure an AC voltage with a minimum of error.
Range
Resolution
Uncertainty (±)
45 Hz à 60 Hz
6 V
0.001 V
2 % + 10 cts
* According to safety rules, 1000V range is limited to 600V.
4.2.4
Frequency (V AC or A AC)
Range
Resolution
Uncertainty (±)
10 to 3000 Hz
0.01 Hz
4.2.5

Duty Cycle

Range
Resolution
Uncertainty (±)
0.1 to 99.9 %
0.1 %
1.2 % + 2 cts
4.2.6
Resistance
Range
Resolution
Uncertainty (±)
600 Ω
0.1 Ω
2 % + 2 cts
6 kΩ
0.001 kΩ
60 kΩ
0.01 kΩ
0.3 % + 4 cts
600 kΩ
0.1 kΩ
6 MΩ
0.001 MΩ
60 MΩ
0.01 MΩ
0.5 % + 20 cts
51 - 138
Sensitivity
0.5 %
15 V RMS
Frequency
5 Hz to 150 kHz

Publicidad

Tabla de contenido
loading

Tabla de contenido