Cleaning and disinfecting

Table 2:
Approved cleaners
Product
b
SaniCloth AF3
PI-Spray II
a
For maximum effectiveness, the component being cleaned must remain wet with disinfectant for a minimum
period of time.
b
Qualified for use as an intermediate-level disinfectant for mycobacteria.
Refer to the cleaners and disinfection document available at
current cleaners and disinfectants.
Table 3:
Approved high-level compatible disinfectants
a
Compatible transducer
Disinfectant
Cidex
rP19xp
Cidex OPA
rP19xp
a
Refer to the cleaners and disinfection document available at
current cleaners and disinfectants.

Safety

WARNING
To avoid device damage or patient injury, do not use the rP19xp needle-guide
bracket on patients with pacemakers or medical electronic implants. The needle-
guide bracket for rP19xp contains a magnet that is used to ensure that the bracket
is correctly oriented on the transducer. The magnetic field in direct proximity to the
pacemaker or medical electronic implant may have an adverse effect.
4
Compatible transducer
rP19xp
rP19xp
www.sonosite.com
Temperature
25
20
www.sonosite.com
Minimum wet contact time
3 minutes
10 minutes
for a complete list of the most
Disinfectant soak duration
o
o
45 minutes
C, 77
F
o
o
12 minutes
C, 68
F
for a complete list of the most
a
Cleaning and disinfecting