7. CLEANING AND DISINFECTION

WARNING: Ensure that the illuminator is de-energized and disconnected from mains power before
attempting to clean and disinfect.
The illuminator can be wiped down with commercially available cleansers commonly used for
disinfection of electronic equipment in hospitals such as ethyl or isopropyl alcohols, disinfecting sprays
containing quaternary ammonium compounds, or hydrogen peroxide.
WARNING: DO NOT use strongly caustic or acidic cleansers such as "Clorox" hypochlorite bleach,
ammonia, muriatic acid, or similar products. DO NOT use acetone, methyl ethyl ketone, or halogenated
/ chlorinated hydrocarbon solvents or cleansers containing any of these restricted compounds.
Apply cleaning agents by light spray or dampened towels. Do not pour liquids onto the device. Do not
allow liquids to enter the device seams or ventilation openings.
Follow all applicable bloodborne pathogen procedures as required by OSHA and/or your hospital
when cleaning and disinfecting the product.
WARNING: The illuminator is not sterilizable. DO NOT attempt to autoclave the device.

7.1 Cleaning the Optic

The optic is located in the topmost port of the rotating turret.
Remove loose dust and surface contaminants on the optic by using compressed air.
Soak a cotton-tipped swab or lens tissue with a solvent mentioned above, then wipe the surface
of the optic in a circular motion starting from the outer edge of the optic.
Repeat as necessary, using a new swab or tissue each time.
8. MAINTENANCE, SERVICING, REPAIR & WARRANTY

8.1 Fuse Replacement

WARNING: Always TURN OFF and UNPLUG the illuminator from mains power before attempting to
replace a fuse.
Turn off illuminator and unplug power cord. On rear panel of the console, remove the fuse holder
located in the appliance inlet. Replace blown fuse with 2A, 5x20mm, fast acting, 250VAC rated
fuses. Insert holder back into fuse housing.

8.2 Warranty

The illuminator carries a 3-year warranty from the date of shipment on workmanship and all defects
of material.
Should your product prove to have such defects within three years of shipment, Sunoptic
®
Technologies
will repair or replace the product or component part without charge. Should your
product(s) need servicing under this warranty, please contact Sunoptic Technologies
distributor for return authorization documentation.
Please carefully pack the unit in a sturdy carton and ship it to the factory. Please include a note
describing the defects, your name, telephone number and a return address. Warranty does not
cover equipment subject to misuse, accidental damage, normal wear, and tear or if transferred to
a new owner without authorization from Sunoptic Technologies
legal rights and you may also have other rights that vary from state to state.
LIT-195 Sunoptic Surgical
Rev. H
®
Page 11 of 104
®
®
. This warranty gives you specific
(English)
or a local

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido