5
4
3
2
1
9
8
7
6
SUB-D 9 (female)
Receive Data
Transmit Data
Data Terminal Ready
System Ground
Data Set Ready & Carrier Detect 1&6
Request to Send
Clear to Send
5
4
3
2
1
9
8
7
6
SUB-D 9 (female)
INTRAsurg
Receive Data
Transmit Data
Data Terminal Ready
System Ground
Data Set Ready & Carrier Detect 1&6
Request to Send
Clear to Send
INTRAsurg
®
500
5
INTRAsurg
PC
SUB-D 9 (female)
PIN
PIN
2
3
Transmit Data
3
2
Receive Data
4
1&6
Data Set Ready & Carrier Detect
5
5
System Ground
4
Data Terminal Ready
7
8
Clear to Send
8
7
Request to Send
13
12
11
10
9
25
24
23
22
21
PC
SUB-D 25 (female)
PIN
PIN
2
2
Transmit Data
3
3
Receive Data
4
6&8
Data Set Ready & Carrier Detect
5
7
System Ground
20
Data Terminal Ready
7
5
Clear to Send
8
4
Request to Send
4
3
2
1
9
8
7
6
8
7
6
5
4
3
2
1
20
19
18
17
16
15
14
DE
M
ONTAGE
B 2.4 Installation PC
Ist der wie in Kapitel B 2.2 be-
schriebene Ja-Zustand sowie die
Schnittstelle auf PC gespeichert, kann der
PC angeschlossen werden. Der PC muß
den technischen Voraussetzungen (siehe
Kapitel A 7) entsprechen.
INTRAsurg 500 und PC ausschalten.
Kabelverbindung herstellen. Der Anschluß
erfolgt am seriellen Port des PCs.
Verwendetes Kabel ist ein handelsübli-
ches Null-Modem Kabel mit beidseiti-
gem handshake (RTS/CTS, DTR/DSR).
Ausführung des Kabels
Anschluß
Anschluß
INTRAsurg
PC
9-polig Buchse
9-polig Buchse**
oder
9-polig Buchse
25-polig Buchse
** Durch entsprechende Adapter kann
auch der Übergang von einer 9-poligen
Buchse auf eine 25-polige Buchse reali-
siert werden.
Maximal zulässige Gesamtlänge
des Null-Modem Kabels: 80 m.
Bei Unklarheiten wenden Sie sich
an Ihren Systembetreuer.
INTRAsurg 500 und PC einschalten.
Die Reihenfolge spielt dabei keine
Rolle.
Zur Ausgabe der Daten auf den PC ein
Terminal Programm starten. In Windows-
Betriebssystemen z.B. das Programm
Hyper Terminal (über Start/Programme/
Zubehör/Hyper Terminal) starten.
B2
113