Specifications

Specifications are subject to change without notice.
Specification
Details
Dimensions (W x D x
13.56 x 8.93 x 12.06 cm (5.34 x 3.52 x 4.75 in.)
H)
Enclosure
304 stainless steel
Light source
Long Life Laser
Weight
0.82 kg (1.8 lb)
Pollution degree
2
Installation category
I
Protection class
III
Power requirements
9-28 VDC (source: Class 2 limited energy, < 150 VA)
Power consumption
Serial and pulse units: 3.3 W; Ethernet unit: 4.3 W;
(maximum)
Analog: 3.5 W; 1 A maximum
Operating
5 to 40 °C (40 to 104 °F); best performance: 10 to 32 °C
temperature
(50 to 90 °F)
Storage temperature
-40 to 70 °C (-40 to 158 °F)
Humidity
Operating and storage: 5 to 95% relative humidity, non-
condensing
Altitude
2000 m (6562 ft) maximum
All manuals and user guides at all-guides.com
Startup
on page 15
diode , Class 3B laser
Specification
Details
Port sizes
Model 6003, 6005: barb fitting for 0.32 cm (1/8-in.) ID
inlet tubing, 0.64 cm (¼-in.) ID outlet tubing
Model 6013, 6015: barb fitting for 0.64 cm (¼-in.) ID inlet
tubing, 0.64 cm (¼-in.) ID outlet tubing
Output signal options
Pulse, analog 4-20 mA, serial RS232 with Modbus RTU
or FXB communication protocol (no networking), serial
RS485 with Modbus RTU or FXB communication
protocol, Ethernet with ModbusTCP protocol
Data storage
1000 samples/records (oldest records are written over
when the buffer is full)
Sample flow rate
Model 6003, 6005: 0.1 cfm (2.83 Lpm) ± 5%
Model 6013, 6015: 1.0 cfm (28.3 Lpm) ± 5%
Inlet pressure
Ambient to 2.5 mm (0.1 in) Hg vacuum
Vacuum requirement
≥ 406 mm (16 in.) Hg (542 mbar) minimum vacuum
measured at each instrument with flow through all
instruments.
Range
Model 6003: 0.3 μm to 10.0 μm at 0.1 cfm (2.83 L/min)
Model 6005: 0.5 μm to 10.0 μm at 0.1 cfm (2.83 L/min)
Model 6013, 6015: 0.5 μm to 10.0 μm at 1.0 cfm
(28.3 L/min)
Sensitivity
Model 6003: 0.3 μm at 0.1 cfm (2.83 L/min)
Model 6005: 0.5 μm at 0.1 cfm (2.83 L/min)
Model 6013: 0.3 μm at 1.0 cfm (28.3 L/min)
Model 6015: 0.5 μm at 1.0 cfm (28.3 L/min)
Counting efficiency
Model 6003: 50% (± 20%) for 0.3 μm, (100% ± 10% at
1.5 times the minimum sensitivity)
Model 6005, 6013, 6015: 50% (± 20%) for 0.5 μm, (100%
± 10% at 1.5 times the minimum sensitivity)
Coincidence loss
Model 6003, 6005 (all output options): 10% at
140,000,000 particles /m
Model 6013, 6015 (all output options except for pulse):
10% at 20,000,000 particles/m
1
.
1
.
3
3
(4,000,000 particles /ft
)
3
3
(566,000/ft
)
English 3

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Met one 6003Met one 6005Met one 6013Met one 6015

Tabla de contenido