List of Error Codes
Error code
Er1
Inflation error
•
When the pressure does not exceed 15
mmHg after operating a pump for 15 seconds
from the start of inflation
•
When the inflation does not reach the set
pressure of the arm cuff within the specified
time after starting the inflation
Er2
Deflation error
•
When the deflation speed is too fast during
the measurement
•
When the deflation speed is too slow during
the measurement
•
When the measurement does not end
within the specified time after starting the
measurement
•
When the pressure is applied more than 5
minutes
Er3
Overpressure error
•
The arm cuff pressure exceeded 299 mmHg.
Er4
Insufficient inflation error
•
Blood pressure could not be measured due
to insufficient inflation after the measurement
ends.
Er5
Indeterminable blood pressure error
•
Blood pressure could not be measured even
when the arm cuff pressure reached the
specified pressure.
Er6
Pulse wave small error
•
Pulse wave was too small.
Er7
Measurement error
•
Relationship between systolic and diastolic
pressures was abnormal.
Er8
Pulse rate error
•
Pule rate did not stay within the range of 30
to 199 beats/min.
Er9
Device error
•
There is a problem with the main unit of the
blood pressure monitor.
22
Cause
•
•
•
•
•
•
•
•
•
•
•
•
•
How to correct
If the connector of rubber tube connecting the
arm cuff and the main unit is loose, connect them
correctly, then measure.
If the rubber tube connecting the arm cuff and the
main unit is bent, stretch it, then measure.
If the arm cuff is wrapped loosely, wrap it correctly,
then measure.
If the rubber tube connecting the arm cuff and the
main unit is bent, stretch it, then measure.
If the measurement is made by setting P-SET to
"AUTO", ask the patient not move his/her arm or
body during the inflation.
When setting the P-SET to "AUTO", turn the Volume
counterclockwise as far as it goes until you can hear
the click sound.
Set the P-SET to 30 to 40 mmHg higher than the
systolic pressure of the patient, then measure.
If the arm cuff is wrapped loosely, wrap it correctly,
then measure.
If the arm cuff is wrapped loosely, wrap it correctly,
then measure.
Do not wrap the arm cuff over a thick cloth or by
rolling up the jacket or shirt.
Check if the patient has an arrhythmia.
Ask the patient not to move his/her arm or body
during the measurement.
Contact the dealer where you purchased the unit or
the nearest OMRON dealer.