5 Pipetting programs

Activity
Designation
Normal pipet-
PIPmode, see PIP
ting
mode, p. 57
Pipetting for
GELmode, see Elec-
electrophoresis
trophoresis (GEL)
mode, p. 61
Mixing samples PIPmix mode, see PIPmix
mode, p. 58
Reverse pipet-
revPIP mode, see revPIP
ting
mode, p. 60
Dispensing
DISPmode, see DISP
mode, p. 63
GEL mode
The GEL mode is not available for Transferpette® -8/-12 electronic 1000 μl and 5000 μl.

5.1 PIP mode

The default program – a previously entered volume is aspirated and dispensed again.
Volume and speed setting, see Set volume, p. 53 and Set the aspiration and dispensing speed, p. 54.
a.
3 x
b.
Transferpette® -8/-12 electronic 997200
Info
Standard program.
A previously entered volume is aspirated and dispensed
again
Program for loading electrophoresis gels. A previously
defined sample volume is aspirated at a high, variable
speed and slowly released again.
Program for mixing liquids.
Sample is aspirated and dispensed repeatedly.
Program especially for pipetting liquids with high viscos-
ity, high vapor pressure, or foaming media.
Program for dispensing liquids.
A volume taken up is dispensed again in partial steps.
NOTICE
a.
Call up menu selection: Pressing the menu key three
times takes you to the program selection.
ð
'Mode' flashes.
b.
Set PIP mode: Use one of the arrow keys to scroll
through the modes until 'PIP' appears.
ð
'Mode' continues to flash.
5 Pipetting programs
57
Operating manual