11 Troubleshooting

11 Troubleshooting
Fault
Instrument does not
respond
Instrument does not
respond
Instrument does not
respond
Instrument does not
respond
Tip dripping/instru-
ment leaking or vol-
ume error
No indication in the
display
Aspiration is not possi-
ble
80
Operating manual
Display
Cause
ERR 1
Rechargeable battery
empty or defective
Electronic components
defective
ERR 2
Electronic components
defective
ERR 3
Unforeseen program
error
ERR 4
No rechargeable bat-
tery in the instrument
Rechargeable battery
defective
Electronic components
defective
Unsuitable tip
Tip not seated tightly
Piston, shaft, or seal
dirty or damaged
Electrostatic discharge Remove rechargeable battery and
Electronic components
defective
Motor has no connec-
tion to the pipetting
unit.
Corrective action
Charge the battery for at least 5
min without actuation. Then con-
tinue working with the charging ca-
ble only until the battery is
recharged. Replace the recharge-
able battery if necessary
Send instrument for repair
Send instrument for repair
Error confirmation by pressing the
'Enter' key, instrument is reinitial-
ized
Insert rechargeable battery
Replace rechargeable battery
Send instrument for repair
Use only high-quality tips
Press tip on more firmly/other in-
terchangeable clip
Clean instrument/replace seal,
grease piston
insert again
Send instrument for repair
Carry out reference run (rEF), see
Reference run (rEF), p. 70.