Commissioning The Gas Warning System - Dräger Pulsar 7000 Serie Instrucciones De Uso

Ocultar thumbs Ver también para Pulsar 7000 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 105
en
Operation
4.2.6

Commissioning the gas warning system

The factory setting of the parameters is described in Chapter 8.4 on Page 67.
The gas warning system can be commissioned in different ways. Any HART
Description (DD) or a computer with Dräger PolySoft can be used for this.
Only the commissioning process with the Emerson Field Communicator 475 HART
display is described below.
NOTICE
Before commissioning, deactivate the alarm activation in the control unit if necessary.
1. Open-path gas detectors only:
Remove the weather protection on the transmitter and receiver.
Duct-mount gas detectors only:
The cover must be removed.
2. Open-path gas detectors only:
Loosen the 8 screws (1, 2, 3, 4) of the gimbal on the transmitter and receiver so that the transmitter and receiver can be
adjusted despite a slight resistance.
3
2
52
1
®
handheld terminal with the Dräger Device
®
handheld terminal and the graphical
4
00200114_00233451

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Pulsar 7000 otr 00 serie

Tabla de contenido