Caution
• Using InBody970 in a dry environment or on a carpet may result in static electricity and damage to the device.
Use an antistatic mat if you need to install in an environment with static device.
• Install the InBody970 on the floor that is flat and vibration-free. If the equipment is installed where the floor is not flat, it may
topple during a test or the test results may be inaccurate.
• Do not clean the foot electrode and the hand electrode with detergent. If liquid cleaner flows into InBody970, it may cause
corrosion and malfunction. For cleaning the InBody970, use an InBody Tissue (wet tissue).
2. Raise the InBody970 upper section to the end and then remove the protective packaging vinyl.
Remove the protective packaging material placed over the footplate.
3. When assembling the hinge cover, push it slightly in the vertical direction so that the spring ball is inserted in the stand end, and
push the hinge cover into the direction of the arrow.
*If the hinge cover is pressed strongly after the spring is inserted fully, or the stand is lowered with it attached, it may cause scratches.
*Be careful not to get your fingers caught when assembling the hinge.
4. Adjust the foot supporters to center the bubble in the level indicator and to level the device for accurate weight
measurement.
*For accurate weight measurement, the device should be leveled.
Caution
•Be careful not to get your hands hurt when handling the foot supporter adjustment screw.
6
Protective
Packaging Vinyl
Protective
Packaging Material
Spring Ball
Foot Supporter Adjustment Screw
Level Indicator
Level Status (O)
Level Status (X)