starlight pro
1.
Place the handpiece into its housing in the charging unit (Figure 5 at page 10 - Ref. B).
The green battery LED will light up (Table 1 at page 11- Ref. B).
2.
the charging phase has been completed when the green battery LED switch off.
5.4
Battery Failure Indication
If the yellow battery LED (check) on the charging unit lights up (Table 1 at page 11- ref. B), this
means that there is a fault with the battery.
NOTE: This failure condition disables operation of the charging unit. To restore proper working
conditions proceed as follows:
1.
Remove the handpiece from the charging unit;
2.
Cut off the power supply to the charging unit for a few seconds (disconnect the power
cable) - All the LEDs will light off;
3.
Reconnect the cable of the charging unit. The green power LED will light up.
5.5
Battery Replacement
to replace a failed battery, contact Mectron Customer Service.
5.6
Battery Safety Requirements
The battery can cause damage to property and/or personal injuries such as burns if conducting
materials such as jewellery, keys or beaded necklaces come into contact with the exposed
terminals.
The conducting material could close an electrical circuit (short circuit) and become very hot.
Make a habit of handling the device with care, particularly if it is placed inside a pocket, bag or
other container in which there are metal objects.
In the case of contact of the terminals with metal objects and consequent short circuit, the lamp
stops and it is necessary to reposition it on the battery charger to resume its operation.
! WARNING: Do not short-circuit the electric contacts of the handpiece with metal
objects or liquids (Figure 6 at page 14 - Ref. A Ref. B).
! WARNING: Do not leave the battery within reach of children.
14
Figure 6 – Electric contacts of the handpiece.
A
B