CAUTION: In case the final user, operating in their own medical room or surgery, in order
to comply with mandatory requirements, must periodically inspect the equipment present
in the surgery, the test procedures to apply to medical electrical equipment and medical
electrical systems for the safety assessment must be carried out following the standard EN
62353 'Medical electrical equipment - Periodic inspections and tests to be carried out after
repair of medical electrical equipment'. the interval for periodic checks, in the intended
operating conditions and described in this "Use and Maintenance" manual, is one year or 2000
hours of use, depending on which of these two conditions occurs first.
WARNING: If an adverse event and/or serious incident attributable to the device occurs
during correct and intended use, it is recommended to report it to the Competent Authority
and to the manufacturer indicated on the product label.