10.1 CANNULA REPLACEMENT

Your nasal cannula should be replaced on a regular basis per the manufacturer's instructions for use. Consult with
your physician and/or equipment provider and/or cannula manufacturer's instructions for replacement information.

10.2 CASE CLEANING

DANGER!
Risk of injury or damage
Liquid will damage the internal components of the concentrator and its equipment. To avoid damage or injury
from electrical shock:
• Turn Off the concentrator and unplug the power cord before cleaning.
• DO NOT allow any cleaning agent to drip inside the air inlet and outlet openings.
• DO NOT spray or apply any cleaning agent directly to the cabinet.
• DO NOT hose down the product.
• DO NOT submerse the device or accessories in liquid
WARNING!
Risk of death, injury or damage
Harsh chemical agents can damage the concentrator and filters
• DO NOT clean with alcohol and alcohol-based products (isopropyl alcohol), concentrated chlorine-based
products (ethylene chloride), and petroleum-based products or any other harsh chemical agents.
Only use mild liquid dish detergent.
Periodically clean the case as follows:
1. Make sure the concentrator is off and is removed from the carry bag.
2. Clean the outside case using a cloth dampened with a mild liquid detergent and water.
3. Allow the concentrator to air dry, or use a dry towel, before returning the concentrator to the carry bag or
backpack and prior to operating the concentrator.
IMPORTANT: The device should receive an external cleaning weekly; accessories should be cleaned as needed.
The device exterior should be cleaned and the output filter replaced prior to delivering to a new patient.
10.3 FILTER CLEANING & REPLACEMENT (RP-501)
The particle filters must be cleaned weekly to ensure the ease of air flow.
To clean:
1. Remove the particle filters from both intake ends of the device.
2. Clean the particle filters with a mild liquid detergent and water, rinse in water and dry fully before reuse.
To purchase additional particle filters, contact your equipment provider or Inogen.
©2022 Inogen, Inc. All rights reserved.