Test strips, control solutions, or sterile lancets may not
be included in the kit (please check the contents on your
product box). They can be purchased separately.
Do a control solution test when:
you suspect the device or test strips are not working
9
properly.
your test results are not consistent with how you feel, or if
9
you think the results are not accurate.
you have dropped or think you may have damaged the
9
device.
To perform the control solution test, do the following:
1. Insert the test strip into the test slot of the device. Wait for
the device to display the test strip "
" ".
2. The meter will detect the difference between control solution
and blood samples automatically. It will automatically mark
the result as a control solution test with "QC" display.
" and blood drop
EN-12
3. Apply the control solution. Shake the control solution vial
thoroughly before use. Squeeze out a drop and wipe it off,
then squeeze another drop and place it on the tip of the
vial cap. Hold the device to move the absorbent hole of the
test strip to touch the drop. Once the confirmation window
is filled completely, the device will begin counting down.
Note:
To avoid contaminating the control solution, do not directly
apply the control solution onto a strip.
4. Read and compare the result. After counting down to 0, the
test result of the control solution will appear on the display.
Compare this result with the range printed on the test strip
vial or individual foil pack and it should fall within this
range. If the test result is out of range, read the instructions
again and repeat the control solution test.
118~160