Preselection Of Running Time; Preselection Of Brake Intensity And Acceleration - OHAUS Frontier FC5706 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

EN-8
The speed is adjustable between 200 rpm and maximum revolution of the centrifuge resp. the maximum
permissible revolution of the pre-selected rotor.
It is the same as the pre-selection of the RCF-value. The setting range is between 20 x g and the maximum
permissible centrifugal force of the rotor.
The maximum speed of the FC5706 is 6000 rpm resp. 4427 x g.
See "Table 3: max. speed and RCF-values for permissible rotor" (See APPENDIX). All important values are
listed there.
ATTENTION:
Please also check the maximum permissible revolutions of your test tubes with the manufacturer.

3.4.2 Preselection of running time

The running time can be pre-selected in three different ranges from 10 seconds up to 99 hours 59 minutes.
1. Range from 10 seconds up to 59 minutes 50 seconds in steps of 10 seconds
2. Range from 1 hour up to 99 hours 59 minutes in steps of 1 minute
3. The continuous run "cont", can be interrupted by the key "Stop"(10).
The running time can be pre-selected with the lid open or closed.
To activate the setting of the running time press the key "Time" (6).
In the display "Time" (A-3) the indication "m : s" or "h : m" flashes, depending on the previous setting.
Set the desired value by using the adjusting knob (1). After exceeding 59 min 50 sec the indication changes
automatically into "h : m". After exceeding 99 hours 59 min the word "cont" appears in the display "Time" (A-3).
That continuous run can only be interrupted by pressing the key "Stop" (10). The time starts counting down as
soon as the set speed is reached.
The display always shows the remaining running time. (See figure 20)
All with number marked passages refer to figure 20
Figure. 20

3.4.3 Preselection of brake intensity and acceleration

This function is activated through the key "Accel/Decel" (5) (refer to figure 21).
By pressing the key once the word "accel" (M7) flashes in the display "Acc/Dec" (A-2). The desired acceleration
can be pre-selected by the adjusting knob (1). The value 0 is equivalent to the lowest and the value 9 to the highest
acceleration.
By pressing the key "Accel/Decel" (5) twice, in the display "Acc/Dec" (A-2), the word "decel"(M8) is shown. Now
the desired brake intensity can be pre-selected by the adjusting knob (1). The value 9 is equivalent to the shortest
and the value 0 to longest possible brake time.
See "table 4: acceleration and deceleration times" (APPENDIX). There the acceleration and deceleration times
for the acceleration and deceleration stages 0 to 9 for permissible rotors are shown.
TM
Frontier
Centrifuge FC5706
A-3
1
10
6

Publicidad

Tabla de contenido
loading