3. SPECIFICATIONS
PARAMETER
Light Source Type
Color Temperature
LED life
CRI
Brightness Control
Light guide adaptor
Input Power
Fuses
Operating Conditions
Storage Conditions
Dimensions
Weight
4. CERTIFICATIONS
PARAMETER
System Classification
Isolation
EMC Certifications
CE Marking
Other Approvals
Degree of protection against
harmful ingress of water
Degree of safety in the
presence of Flammable
Anesthetics
Mode of operation
LIT-216 Sunoptic Surgical
Rev. E
VALUE
LED (Light Emitting Diode)
5600K nominal for High CRI models
6500K nominal for Standard CRI models
30,000 hours (typ.)
75 nominal for Standard CRI models
92 nominal for High CRI models
PWM (Pulse-width Modulation) 0-100% dimming
Rotating turret with STORZ, ACMI, WOLF, and OLYMPUS (if equipped)
100-240 VAC, 50/60 Hz
0.7 – 1.5A (max.)
5 x 20 mm, 250V, 2A, Type F
32 to 104⁰F (0 to 30⁰C), 30 to 85% RH non-condensing, 700 to 1060
hPa
-4 to 140⁰F (-20 to 60⁰C), 0 to 95% RH non-condensing, 700 to 1060
hPa
11.2" x 4.5" x 13.2" (W x H x D)
28.5 x 11.4 x 33.6 cm (W x H x D)
7.55 lbs / 3.42 kg
VALUE
FDA Class I, Device Listing D095692, 510(k) exempt
EU Class I, Active device per Annex IX, rule 1
Type BF
CISPR 11 Class A, IEC 60601-1-2 4
Electrostatic discharge: ±8 kV contact, ±15 kV air
Radiated RF EM Fields: 3 V/m, 80 – 2700 MHz
EFT / Burst: ±2 kV, ±1 kV signal lines, 100 kHz
SURGE: ±0.5, ±1 kV
Conducted disturbance: 3 V 150 kHz – 80 MHz and 6 V in ISM
bands
Power frequency magnetic fields: 30 A/m
Voltage dips: 0% Un / 0.5 cycles at 0°, 45°, 90°, 135°, 180°, 225°,
270°, 315°, 40 % Un / 5 cycles, 70 % Un / 25 cycles
Voltage interruptions: 0% for 5000 ms
Proximity fields: in accordance with EN 60601-1-2: 2015 table 9
Regulation (EU) 2017/745
CAN/CSA-C22.2 No. 60601-1-08, TC 2:2011 (Corrigendum 2)
CAN/CSA-C22.2 No. 60601-2-18:11
IPX-0; no protection.
Equipment is NOT suitable for use in the presence of flammable
anesthetics.
Continuous
®
Page 6 of 114
th
Edition
(English)

Publicidad