Installation and connections
The pulse oximeter is delivered with the following components: a practical imitation leather case, 4
"AA" alkaline batteries, AC adapter with 6 V output, 600 mmA, a finger sensor for heart beat survey.
To feed the equipment with 4 "AA" batteries, open the apposite compartment on the back and insert the
batteries. Pay attention not to invert the direction of poles (see the picture inside the battery compartment)
and then replace the cover.
To feed the equipment with an AC adapter, insert the adapter jack into the corresponding socket on the
equipment base and plug in the adapter.
With the exception of rechargeable batteries, remove common or alkaline batteries when
using the AC adapter for charging.
Insert the finger sensor connector into the pulse oximeter (when the sensor is not inserted, the display
views the following writing: Prb). It is sufficient to make the grooves present on the connector and on
the socket match.
Instrument description and symbols
PWR
Pilot light Equipment ON.
Select
Pushbutton for the selection of SpO 2 Max and Min values and PULSE/min.
Pushbutton to increase SpO 2 Max and Min values and PULSE/min.
Pushbutton to decrease SpO 2 Max and Min values and PULSE/min. Or ENTER for some
functions.
VOL.
To select the level of sound volume, press this button.
Mute
Turn on/off audio signals
Run-down battery indicator. When the battery charge is low, the low battery indicator is
GREEN. When the GREEN light is on, the equipment can work for another half-hour/an
hour. It is time to replace the batteries.
Under low-battery conditions, printing function is immediately disabled.
Memory Indicator for memory function enabled.
Timer
Indicator for timer function enabled.
Prb
Sensor not connected with the equipment
Pot
No patient connected
Lo
Low signal
PT
Printing pushbutton. Press this button to print the last 6 data stored in memory. This button
can be used only when the battery charge level is good
FD
Paper sheet forward feed button. Press this button to make the sheet advance a line at a time.
This button can be used only when the battery charge level is good
INSTALLATION
OPERATION
11