Measurements
11. Measurements
11.1 Basic Functions
11.1.1 Factor
11.1.2 Standard
Pressing the
icon will open the basic functions menu with six
different measurement options to choose from. With the excep-
tion of the turbidity measurement, the tests in the Basic Functions
menu are independent of the cuvette size used.
After a measurement, various sample information can be entered
via the icons below the results window.
Using the icon
, further sample information can be added below
the measurement result. For this purpose the desired sample in-
formation can be activated using the slider in front of the option.
Further options are available via the
With the command Do not save the measurement result is not
saved in the instrument memory when the cuvette is pulled out or
when leaving the menu.
This function is only available in Open Mode.
The basic function Factor enables the determination of measure-
ment results through the multiplication of an absorbance with a
user defined factor.
After having selected this method, a window will appear asking for
the factor (max. 7 digits) and the wavelength for the measurement
and calculation.
Confirming with
opens the measurement window. Inserting the
zero solution and pressing
After the successful measurement of the zero solution, the sample
solution needs to be inserted. Pressing
ment process. After the measurement has been completed, the
measurement result will be displayed in the result window.
The basic function Standard enables the definition of a sample's
concentration on basis of the concentration of a standard solution
and a zero solution. After having selected this method, a window
will appear asking for the standard concentration (max. 7 digits)
and the wavelength for the measurement.
Confirming the entries by pressing
window. Inserting the zero solution and pressing
the zero measurement. Then, the standard solution needs to be
inserted. Pressing
Spectrophotometer NANOCOLOR
icon.
NOTICE
will initiate the zero measurement.
will start the measure-
will open the measurement
will start the sample measurement.
Advance 1.0, 11.2020
®
will initiate
17