QUICK GUIDE ____________________________________________________________________
NOTE
NOTE
NOTE
NOTE
10 __________________________________________________________________ M211281EN-B
8.
Insert the probe into a known reference temperature (if HMK15
Humidity Calibrator is not used) and let the temperature reading
stabilize.
Do not touch the adjustment buttons before the conditions have
stabilized.
9.
Using the - and + buttons, make the temperature offset adjustment
by making sure the U
- or + at least once even if the value is correct) and press the ADJ
button. The red indicator LED begins to blink faster (400ms
cycle time).
If you do not wish to perform the two-point T adjustment at this
time, press the ADJ button one more time to finish the calibration,
then skip the rest of the steps.
10.
Insert the probe into another reference temperature.
Do not touch the adjustment buttons before the conditions have
stabilized.
11.
Using the - and + buttons, make the temperature gain adjustment
by making sure the U
either - or + at least once even if the value is correct).
12.
Press the ADJ button one more time. The red indicator LED turns
off and the transmitter returns to normal mode. The calibration
procedure is now finished.
In case of calibration error both LEDs are blinked alternately at very fast
rate (cycle time 200ms) for a period of 2s after which the transmitter
returns to normal mode.
In case of using a HMT130 transmitter with display option, the
following texts are shown at the display during calibration:
"Probe cal: RH 1" corresponding the green LED blinking slowly
"Probe cal: RH 2" corresponding the green LED blinking fast
"Probe cal: T 1" corresponding the red LED blinking slowly
"Probe cal: T 2" corresponding the red LED blinking fast
"Probe cal: Error" corresponding both LEDs blinking alternately
at very fast rate
voltage is correct (you have to press either
out
voltage is correct (you have to press