Specifications - WIKA PG43SA-S Manual De Instrucciones

Ocultar thumbs Ver también para PG43SA-S:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

8. Dismounting and disposal / 9. Specifications
8.1 Dismounting
Only disconnect the measuring instrument once the system has
been depressurised!
8.2 Disposal
Incorrect disposal can put the environment at risk.
Dispose of instrument components and packaging materials in
an environmentally compatible way and in accordance with the
country-specific waste disposal regulations.

9. Specifications

9.1 Pressure limitation
Model
Steady
PG43SA-S Full scale value
PG43SA-C 3/4 x full scale value 2/3 x full scale value
9.2 Overload safety
Model
Overload safety
PG43SA-S 2 x full scale value, max. 40 bar or max. pressure rating
(PN) of the process connection
PG43SA-C 5 x full scale value, max. 40 bar or max. pressure rating
(PN) of the process connection
1) see PN in the tables of data sheets PM 04.15, PM 04.16
9.3 Vacuum safety (only model PG43SA-S)
All scale ranges are suitable for the vacuum typical of CIP or SIP
cleaning. Positive scale ranges < 6 bar (< 100 psi) are optionally
available in a vacuum-tight version (to -1 bar).
WIKA operating instructions pressure gauge, PG43SA-S, PG43SA-C
Fluctuating
Scale range
< 4 bar
Full scale value
1)
Scale range
≥ 4 bar
2/3 x full scale
value