System start up
•
LCK 360: 0.2-6 mg/l solved Zinc
•
LCK 304: 0.24-7.2 mg/l total Zinc after digestion with Crack-Set LCW 902
Each programming procedure has only to be followed once prior to the first test
evaluation.
For detailed information please read the POCKET COLORIMETER II instruction
manual.
3.2.2
Data table
3.2.2.1
Evaluation as Zinc
Table 1 Evaluation as solved Zinc and total Zinc
RES (resolution)
S0 (standard 0)
A0 (absorbance 0)
S1 (standard 1)
A1 (absorbance 1)
3.3 Programming procedure
Note: Each programming procedure has only to be followed once prior to the first test
evaluation.
Note: After programming the calibration data, it could be helpful to mark the channel(s) 1
and/or 2 accordingly with a label.
3.3.1
LCK 360 solved Zink Zn
1.
Press POWER to switch on the instrument.
2.
Press MENU. "SEL" is displayed. A flashing arrow indicates the current
measuring range (1 or 2).
3.
Select measuring range 1 for LCK 360 (0.2-6 mg/l Zn).
GREEN
Note: Press
4.
Press MENU for about five seconds until "USER" followed by "CAL" is
displayed.
5.
Press BLUE key until "Edit" is displayed. Press GREEN key. " RES" is
displayed.
6.
Press GREEN key. "-.---" is displayed.
7.
Press BLUE key until "--.--" is displayed. Press GREEN key. "S0" is displayed.
22
LCK 360
solved Zinc
--.--
00.20
0.035
06.00
0.610
key to toggle between measuring ranges.
LCK 360
total Zinc
--.--
00.24
0.035
07.20
0.610