System start up
8.
Press GREEN key until "A0" is displayed.
9.
Press GREEN key until "S1" is displayed.
10. Press GREEN key. Enter "06.00" by using BLUE key to change value and
GREEN key to move to the next space. Press GREEN key until "A1" is
displayed.
11. Press GREEN key. Enter "0.610" by using BLUE key to change value and
GREEN key to move to the next space. Press GREEN key until "Add" is
displayed.
12. Press MENU until "Edit" is displayed. Press GREEN key until „S0" is displayed.
13. Press GREEN key. Enter "00.20" by using BLUE key to change value and
GREEN key to move to the next space. Press GREEN key until "A0" is
displayed.
14. Press GREEN key. Enter "0.035" by using BLUE key to change value and
GREEN key to move to the next space. Press GREEN key until „S1" is
displayed.
15. Press MENU until "Edit" is displayed to check the entered data. Press GREEN
key. „RES" is displayed.
16. Press BLUE key to scroll through the entries. Compare the entries with the data
shown in the table below.
17. After comparing the entries press MENU two times to enter the measuring
mode. See
4.3 Evaluate LANGE Zinc Cuvette Test on page
Note: In case of any mistake retrieve the factory calibration first. See
calibration on page
again. If the instrument is shut-off or loses power during data entry, all edits will be lost.
Automatic shut-off is 60 minutes.
3.3.2
LCK 360 total Zinc Zn
1.
Press POWER to switch on the instrument.
2.
Press MENU. "SEL" is displayed. A flashing arrow indicates the current
measuring range (1 or 2).
3.
Select measuring range 2 for LCK 360 (0.24-7.2 mg/l Zn).
GREEN
Note: Press
4.
Press MENU for about five seconds until "USER" followed by "CAL" is
displayed.
5.
Press BLUE key until "Edit" is displayed. Press GREEN key. " RES" is
displayed.
Table 2 Calibration data of LCK 360 solved Zn
RES
S0
A0
S1
A1
25. After retrieving the factory calibration, start the programming procedure
key to toggle between measuring ranges.
--.--
0.20
0.035
6.00
0.610
27.
3.5.1 Retrieve the factory
23