ERROR INdICATORS ANd TROUBLESHOOTING TIPS
ERROR INDICATORS
SyMBOL
Air.plug.disconnected .
Arm.cuff.not.applied.correctly .
Air.is.leaking.from.the.arm.cuff .
Cuff.over.or.under.inflated .
The.arm.cuff.was.inflated.above.299.
mmHg.when.inflating.the.cuff.manually .
Movement.during.measurement .
Clothing.is.interfering.with.the.
arm.cuff .
Device.error .
Sensor.may.be.out.of.alignment .
Cuff.is.not.applied.correctly .
CAUSE
47
CORRECTION
Insert.the.plug.securely ..Refer.to."Applying.the.
Arm.Cuff" .
Apply.the.arm.cuff.correctly ..Refer.to."Applying.
the.Arm.Cuff" .
Replace.the.cuff.with.the.new.one .
Repeat.measurement ..Remain.still.and.do.not.
talk.during.measurement ..Refer.to."Taking.a.
Measurement" .
If."E2".appears.repeatedly,.inflate.the.cuff.manually.
until.it.is.30.to.40.mmHg.above.your.previous.
measurement.result ..Refer.to."Taking.a.Measurement" .
Do.not.inflate.the.cuff.above.299.mmHg ..Refer.to.
"Taking.a.Measurement" .
Repeat.measurement ..Remain.still.and.do.not.
talk.during.measurement ..Refer.to."Taking.a.
Measurement" .
Remove.any.clothing.interfering.with.the.arm.cuff ..
Refer.to."Applying.the.Arm.Cuff" .
Contact.Customer.Service .
Contact.Customer.Service.for.repair.information .
Refer.to."Applying..the.Arm.Cuff" .