ART42490-004_rev-A 10:30am 04-May-22
Electromagnetic environment – guidance: The Sensor is suitable for use in all
establishments, including domestic establishments and those directly connected
to the public low voltage power supply network that supplies buildings used for
domestic purposes.
Guidance and manufacturer's declaration – electromagnetic
immunity
The Sensor is intended for use in the electromagnetic environment specified
below. The customer or the user of the Sensor should assure that it is used in such
an environment.
Immunity test: Electrostatic discharge (ESD); IEC 61000-4-2
IEC 60601 test level: ± 8kV contact; ± 2 kV, 4 kV, 8 kV, 15 kV air
Compliance level: ± 8kV contact; ± 2 kV, 4 kV, 8 kV, 15 kV air
Electromagnetic environment – guidance: Floors should be wood, concrete or
ceramic tile. If floors are covered with synthetic material, the relative humidity
should be at least 30%.
Immunity test: Power frequency (50/60 Hz); magnetic field; IEC 61000-4-8
IEC 60601 test level: 30 A/m 50 Hz or 60 Hz
Compliance level: 30 A/m 50 Hz or 60 Hz
Electromagnetic environment – guidance: Power frequency magnetic fields should
be at levels characteristic of a typical location in a typical domestic, commercial or
hospital environment.
Immunity test: Radiated RF; IEC 61000-4-3
IEC 60601 test level: 10 V/m; 80 MHz to 2.7 GHz; 80% AM at 1 KHz
Compliance level: 10 V/m; 80 MHz to 2.7 GHz; 80% AM at 1 KHz
Electromagnetic environment – guidance: