EN-36
7
MAINTENANCE
CAUTION: DISCONNECT THE UNIT FROM THE POWER SUPPLY BEFORE CLEANING.
7.1
Model T31P Cleaning
• The housing may be cleaned with a cloth dampened with a mild detergent if necessary.
• Do not use solvents, chemicals, alcohol, ammonia or abrasives to clean the housing or control panel.
7.2
Model T31XW Cleaning
• Use approved cleaning solutions for the stainless-steel Indicator housing and rinse with water. Dry thoroughly.
• Do not use solvents, chemicals, alcohol, ammonia or abrasives to clean the control panel.
7.3
Troubleshooting
SYMPTOM
Unit will not turn on.
Cannot zero the Scale, or will not zero when
turned on.
Unable to calibrate.
Cannot display weight in desired weighing
unit.
Cannot change menu settings.
Battery indicator is flashing. (T31P Only)
Battery fails to charge fully. (T31P Only)
Error 7.0
TABLE 7-1. TROUBLESHOOTING.
PROBABLE CAUSE(s)
Power cord not plugged in or properly
connected.
Power outlet not supplying electricity.
Battery power used up. (T31P Only)
Other failure.
Load on Scale exceeds allowable limits.
Load on Scale is not stable.
Load Cell damage.
Lock Calibration Menu set to On.
Lock switch is "on".
LFT menu set to On.
Incorrect value for calibration mass.
Unit not set to On.
Menu has been locked.
Lock switch set on.
Battery discharged.
Battery is defective.
Unstable weight reading when defining
reference weight.
3000 Series Indicators
REMEDY
Check power cord connections. Make sure
power cord is plugged in properly into the
power outlet.
Check power source.
Reconnect AC power to charge the battery.
Service required.
Remove load on Scale.
Wait for load to become stable.
Service required.
Set Lock Calibration Menu to Off.
Refer to Section 3.9 Menu Lock.
Set the Lock switch to Off.
Set LFT menu to Off.
Use correct calibration mass.
Enable unit in the Units Menu.
Refer to Section 3.7 in the Unit Menu.
Set selected menu to Off in the Lock Menu.
Lock Switch on the circuit board may
need to be set to the Off position.
Set the Lock switch to off.
Connect indicator to power and charge
battery.
Have the battery replaced by an authorized
Ohaus service dealer.
Unstable Error, check platform location.