Capacitance measurement
Range
60.00 nF
600.0 nF
6.000 µF
60.00 µF
600.0 µF
6.000 mF
60.00 mF
100.0 mF
Overload protection: 300 V DC/AC RMS
Frequency measurement
Range
9.999 Hz
99.99 Hz
999.9 Hz
9.999 kHz
99.99 kHz
Sensitivity: >5 V RMS for 20...80% of duty cycle and <100 kHz
Overload protection: 1000 V DC/AC RMS
Duty cycle measurement
Range
10.0...90.0%
Pulse amplitude: ≥5 V
Pulse width: 0.1 ms...100 ms
Frequency: 40 Hz...10 kHz
Temperature measurement
Range
-20.0...+1000C
-4.0...+1832F
The accuracy of the temperature probe is not taken into account
Overload protection: 300 V DC/AC RMS
66
Resolution
0.01 nF
0.1 nF
0.001 µF
0.01 µF
0.1 µF
0.001 mF
0.01 mF
0.1 mF
Resolution
0.001 Hz
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
Resolution
0.1%
Resolution
0.1 or 1C
0.1 or 1F
CMP-1010 – USER MANUAL
Accuracy
(4.0% m.v. + 20 digits)
(3.0% m.v. + 5 digits)
(5.0% m.v. + 5 digits)
(5.0% m.v. + 8 digits)
(5.0% m.v. + 15 digits)
Accuracy
(1.2% m.v. + 5 digits)
Accuracy
(1.2% m.v. + 2 digits)
Accuracy
± (3% m.v. + 5C)
± (3% m.v. + 9F)