Table 2
Guidance and manufacturer's declaration-electromagnetic emission
The Handheld Pulse Oximeter is intended for use in the electromagnetic
environment specified below. The customer or the user of the Handheld Pulse
Oximeter should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge(ESD)
IEC61000-4-2
Electrical fast
transient/
burst
IEC61000-4-4
Surge
IEC 61000-4-5
III EMC Compliance - ENGLISH
IEC60601 test
level
±8 kV contact
±2 kV, ±4 kV,
±8 kV, ± 15
kV air
±2kV for
power
Supply lines
±1kV for Input
a.c. Power
Ports
±0.5 kV, 1kV
line (s) to
line(s)
±0.5 kV, ±
1 kV, ±2kV
line(s) to earth
Compliance
level
±8 kV contact
±2 kV, ±4 kV,
±8 kV, ± 15
kV air
±2kV for
power
Supply lines
±1kV for Input
a.c. Power
Ports
±0.5 kV, 1kV
line (s) to
line(s)
±0.5 kV, ±
1 kV, ±2kV
line(s) to earth
Electromagnetic environ-
ment -guidance
Floors should be wood,
concrete or ceramic tile.
if floors are covered with
synthetic material, the
relative humidity should
be at least 30%
N/A
N/A
148