9 Troubleshooting
Trouble
Unstable SpO
2
and Pulse Rate
display
Unable
to measure
Temperature
Device will not
switch on
No Display
No Signal
Possible Reason
1. The finger is not placed
far enough inside.
2. The finger is shaking
or the patient is moving.
1. Temperature probe
is not connected properly
1. The batteries are drained
or almost drained.
2. The device is
malfunctioning.
1. The device will power off
automatically when there is
no signal and no operation
for 1 minute.
2. The battery voltage is low.
1. Probe off or incorrect
connection.
2. Incorrect finger insert
3. Probe is damaged.
ENGLISH
Solution
1. Place the finger
correctly inside and try
again.
2. Reduce patient
movement.
2. Reinsert the probe into
the device
1. Recharge battery.
2. Please contact the local
service center.
1. Normal.
2. Recharge battery.
1. Reconnect the probe.
2. Reinsert the finger.
3. Replace a new probe.