Figure 11: Measurement function U
1 Display of measurement function, parameters and set limit value
2 "
t /
F:
(Residual voltage or Discharge time)
3 "MODE" menu key, for selection of measurement mode "LIN"
or "UNLIN"
4 "TIME" menu key, for selection of measurement time (1 s or
5 s)
t
RES
U
F:
2
Fig 12: Measurement function t
1 Display of measurement function, parameters and limit value
2 "
t /
F:
(Residual voltage or Discharge time)
3 "LIM" menu key, for selection of discharge time limit value (1 s
or 5 s)
• For saving displayed measurement result press the "SAVE" function
(12) twice, for further instructions see the "Memorizing example"
section.
• If in linear mode the mains voltage is disconnected at too low value
(<20% of peak value), it means scaling of measurement result to peak
value is not possible. In this case "REPEAT THE MEASUREMENT" message
appears on the display and the measurement must be repeated.