Descargar Imprimir esta página

Beha-Amprobe MT204-S Manual Del Usuario página 42

Ocultar thumbs Ver también para MT204-S:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Fuse replacement (mains fuse F1)
In case mains switch pilot lamp (3) does not illuminate after connecting the
Machinery-Tester MT204-S to mains outlet and switching on the mains
switch and neither the LC-Display (16) shows any indication, it is very likely
mains fuse (2) to be blown.
To replace the fuse proceed as follows:
1)
Open the fuse holder 1 (2) above the ON/OFF switch by using an
appropriate screwdriver
2)
Remove the defective fuse and replace it with a new one (T 1 A /
250 V, 5 × 20 mm)
3)
Close the fuse-holder again
Fuse replacement (fuse F2 for R
functions)
Internal fuse F2 (FF 12.5 A / 500 V, 6.3 × 32 mm) has blown if:
• Text "FUSE (F2) 12.5A" appears on the LC-Display in R
function
• Measurement values in Z
I
higher than 4.6 kA)
PSC
• Check the fuse also if "FUSE (F2/F3)" appears on the LC-Display in R
or R
function
PE10A
To replace the fuse proceed as follows:
1) Unlock corresponding fuse holder (4) by using an appropriate
screwdriver
2)
Remove the defective fuse and replace it with a new one (FF 12.5 A
/ 500 V, 6.3 × 32
3) Lock the fuse holder again.
PE10A
function are very low (Z
I
mm).
42
, Z
, R
and HV
I
ISO
, R
PE10A
ISO
lower than 0.05 Ω or
I
or HV
PE0.2A

Publicidad

loading

Este manual también es adecuado para:

3454632