8.3
Waveform Outputs
Analogue
Digital
CAL function
8.4
General
Max. Audio Output
Auto shut-off
Headphone output
RS232 Interface
(MD2 only)
Battery Type and Life
Size
8.5
Environmental
Operating
Temperature range
Relative Humidity
Pressure
Transport and Storage between uses
Without relative humidity control
At a r.h. of up to 90% non-condensing
At a water vapour pressure up to 50 hPa
8.6
Standards Compliance
IEC 60601-1: 2012
IEC 60601-2-37: 2015 Thermal Indices (TI) and Mechanical Index
(MI) are below 1.0 for all device settings.
8.7
Accessories
Use only the recommended accessories. See www.huntleigh-diagnostics.com for a list of accessories.
Zero crosser, 3.5 V full scale per channel (forward and reverse). Conversion
factor automatically adjusted to give full scale outputs at ± 0.1% (± 10%) of probe
frequency (e.g. ± 8 kHz for VP8)
Data formatted to interface with Recorder software running on compatible computer.
Conversion factors automatically set as per analogue outputs.
Automatic sequence of CAL levels set to ± 0.05 % (± 10 %) and ± 0.1 % (± 10 %)
of probe frequency (e.g. ± 4 kHz & ± 8 kHz for VP8) with zero baseline at start and
end of sequence.
500mW rms typical (loudspeaker)
After three minutes of no signal or ten minutes of use
Max. output Power:
25 mW rms (32 Ω)
Connector:
3.5 mm stereo jack socket
Max. applied voltage: +9 Vdc
Data format: RS232
Connector: 8 pin subminiature DIN socket
Max. applied voltage: +5 Vdc
IEC 6LR61 or IEC 6LP3146, Typically 250 one minute examinations
140 x 27 x 74 mm
Weight
+5°C to +40°C
15% to 90% (non condensing)
700 hPa to 1060 hPa
-25°C to +5°C
+5°C to +35°C
>+35°C to +70°C
IEC 60601-1-11: 2015
IEC 60601-1-2: 2014
295 g