Safety Instructions
Modification or opening of the Electrosurgical Generator by unqualified persons may result in
electrical shock and will violate the warranty of the Electrosurgical Generator.
In the event of failure of the Electrosurgical Generator, there is a risk of undesirable rise in
output power that might cause unintended tissue effects. To minimize possible failure, have
the Electrosurgical Generator calibrated and checked for safety once a year.
In the event of a malfunction or failure, immediately cease treatment and switch the Electro-
surgical Generator OFF using the On / Off Power switch on the Rear Panel or by disconnecting
the power supply cord from the Electrosurgical Generator. In emergencies, disconnecting the
Instrument from the Interface Cable may also terminate treatment and energy delivery.
To avoid risk of electric shock the Electrosurgical Generator must be connected via its AC
power cord to mains supply with a protective earth.
To avoid risk of fire replace fuses only as marked.
USA only: Grounding reliability can only be achieved when the equipment receptacle is
marked "Hospital Grade".
RFID emitters/readers transmit signals that might cause electromagnetic interference that
could disrupt the operation of the Creo Electrosurgical System. Keep all RFID equipment at
least 1 m away from this device system. Be aware that RFID emitters can be portable or hid-
den in the room or building structure.