Cleaning
Caution
Cleaning does not render the Electrosurgical Generator surfaces sterile.
Disconnect the Electrosurgical Generator from the AC supply before cleaning to avoid risk of
electrical shock.
Completely remove and allow all cleaning liquids to dry before connecting the Electrosurgical
Generator to the AC supply to avoid risk of electrical shock.
The Electrosurgical Generator, AC power cord and Footswitch should be cleaned before and after each
use in accordance with procedures or other requirements of the hospital or other clinical site.
The external surfaces should be cleaned only by wiping with a clean, non-abrasive disinfecting wipe that
is lightly dampened with the following:
70/30 Isopropyl Alcohol (IPA)/Water mix.
In the event of any damage or degradation occurring to the Electrosurgical Generator as a result of
cleaning, please contact Creo Medical immediately for advice. Do not use the Electrosurgical Generator
when it appears to be damaged.
Refer to the Instructions for Use of the Footswitch for cleaning.
Proprietary & Confidential - Page 31 of 391 - Uncontrolled if Printed
28