Always place the Footswitch near the feet of the user such that deliberate action by the
STEP 9.
user is required to activate the intended output from the Electrosurgical Generator.
If needed, the Creo Electrosurgical Generator can be connected to the external grounding system of the
room via the standard Equipotential Terminal Pin (15) on the back of the Generator using a connecting
cable designed for this purpose. External grounding removes the hazards presented by low frequency
leakage currents in the instance of a defective grounding system within the room.