Hinweise zur elektromagnetischen
Verträglichkeit (EMV)
Guidance and manufacturer's declaration – electromagnetic immunity
HALOGEN 250 twin model 20 1133 20 is intended for use in the electromagnetic environment specified below.
The customer or user of the HALOGEN 250 twin model 20 1133 20 should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic Discharge (ESD)
± 6 kV contact
IEC 61000-4-2
± 8 kV air
Electrical fast
± 2 kV for power supply lines
transient/burst
± 1 kV for input/output lines
IEC 61000-4-4
Surge
± 1 kV differential mode
IEC 61000-4-5
± 2 kV common mode
Voltage dips, short
<5 % U
interruptions and voltage
(>95 % dip in U
variations on power supply
for 0.5 cycle
input lines
IEC 61000-4-11
40 % U
(60 % dip in U
for 5 cycles
70 % U
(30 % dip in U
for 25 cycles
<5 % U
(>95 % dip in U
for 5 sec
Power frequency
(50/60 Hz)
3 A/m
magnetic field
IEC 61000-4-8
* Note: U
is the a.c. mains voltage prior to application of the test level.
T
Appendix
Electromagnetic Compatibility
(EMC) Information
Table 202
Compliance level
Complies
± 6 kV contact
± 8 kV air
Complies
± 2 kV for power supply
lines
± 1 kV for input/output lines
Complies
± 1 kV differential mode
± 2 kV common mode
Complies
*
<5 % U
*
T
)
T
(>95 % dip in U
T
for 0.5 cycle
40 % U
T
(60 % dip in U
)
T
)
T
for 0.5 cycle
T
See guidelines
70 % U
T
(30 % dip in U
)
T
)
T
for 1 cycle
T
See guidelines
Complies
<5 % U
T
)
T
(>95 % dip in U
T
for 5 sec
Complies
3 A/m
Anexo
Indicaciones sobre compatibilidad
electromagnética (CEM)
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
)
T
If the user of the equipment or system requires
continued operation during power cuts, short term
power interruptions or if power supply is erratic, it
is recommended that the equipment or system be
powered from an uninterruptible power supply.
)
T
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.