For maintenance, use only specified spare parts. The
manufacturer will not be held responsible for any accident
occuring following a repair done other than by its After
Sales Service or approved repairers.
5.1

Servicing

When the
batteries of the CA 6460 or recharge the battery of the CA
6462. Check that no other terminals are connected before
opening the equipment.
5.1.1
Replacing the batteries or rechargeable cells
I Detach the four captive screws under the housing
I Then remove the shell + front panel assembly of the yellow
housing
I Then undo the two screws closing the battery compartment
cover
I Take out the 8 cells and replace them
Note: It is possible to replace the batteries by rechargeable cells
(1.2 V - 2 Ah or above, NiMH, of same size):
I Remove the plug under the batteries
I Set the switch to the position: NiMH
I Replace the plug.
I Insert the 8 rechargeable cells
Then in both cases:
I Replace the battery cover
I Undo the two screws closing the battery compartment
cover.
I Replace the shell + front panel assembly in the yellow
housing then tighten the four captive screws under the
housing
5.1.2

Recharging all replacing the battery (C.A. 6462)

I Connect the battery charge connector to the mains
I The CHARGE indicator light comes on steadily in red
I When the battery is charged, the CHARGE indicator light
comes on steadily in green
I The charge time is approximately six hours for charging to
80% of the battery capacity. It is possible to top up the
charge.