Oscillo-Saw GOLD II
4.2 Contraindication
There are no known contraindications.
5 Mounting / exchanging the saw blade
Danger of injury due to unintentional activation. Pull out the plug before
changing the saw blade. If applicable use thick gloves to prevent injury
when handling the saw blade. Only ever connect the HEBU oscillating
cast cutting saw to the mains when it is switched off.
5.1 Releasing the retaining screw / removing the old saw blade
1. Place the cast cutting saw on a stable underlay (table).
2. Apply one fork wrench at the drive shaft behind the saw blade,
and the other one at the front at the retaining screw (see fig. 1).
3. Hold the rear wrench firmly and use it to fix the cast cutting saw.
4. Release the retaining screw using the front wrench.
5. Remove the retaining screw and the fixing ring.
Ensure that you do not lose any components.
6. If applicable, remove the old saw blade.
5.2 Inserting the new saw blade
1. Check that the female thread and saw blade support of the drive shaft are clean.
2. Position the new saw blade.
It makes no difference which side of the saw blade is facing inwards or outwards.
Important!
The notches in the saw blade must be positioned precisely flush on the spigots of the
drive shaft.
Fig. 1:
Releasing the retaining screw
EN
21