arjo Skin IQ Instrucciones De Uso página 15

Microclimate manager
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 84
Guidance and manufacturer's declaration - electromagnetic immunity
The Skin IQ MCM is intended for use in the electromagnetic environment specified below. The customer or user of the Skin
IQ MCM should assure that it is used in such an environment.
Immunity Test
Electrostatic
±8 kV Contact
discharge (ESD)
IEC 61000-4-2
Electrical fast
±2 kV for power
transient / burst
input/output lines
IEC 61000-4-4
Surge
1 kV line(s) to Line(s)
IEC 61000-4-5
2 kV line(s) to Earth
5% half cycle
Voltage dips, short
interruptions and
40% for 5 cycle
voltage variations on
70% for 25 cycle
power supply input
lines
IEC 61000-4-11
5% for 5 seconds
Power frequency
(50Hz/60Hz)
magnetic field
IEC 61000-4-8
NOTE: U, is the a.c. mains voltage prior to application of the test level.
Recommended separation distances between portable and mobile RF communications
equipment and the Skin IQ MCM
The Skin IQ MCM is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Skin IQ MCM can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the Skin IQ MCM as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum
output power
of transmitter
W
150 kHz to 80 MHz
not applicable
0.01
N/A
0.1
N/A
1
N/A
10
N/A
100
N/A
For transmitters rated at a maximum output power not listed above, the recommended separate distance d in metres
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1, At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2, These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from surfaces, objects and people.
IEC 60601
Compliance Level
Test Level
±8 kV Contact
±15 kV Air
supply lines
±1 kV lines to ±2 kV
for power supply lines
±1 kV for
1 kV line(s) to Line(s)
2 kV line(s) to Earth
5% half cycle
40% 5 cycles
70% 25 cycles
5% for 5 seconds
30 A/M
Separation distance according to frequency of transmitter
80 MHz to 800
MHz
3.5
P
d=[ — ]
E
1
0.12
0.37
1.2
3.7
12
Electromagnetic Environment Guidance
In accordance with IEC 60601-1-2, if
the floors are covered with synthetic
material, the relative humidity should be
±15 kV Air
at least 30%
Power frequency magnetic fields should
be at levels characteristic of a typical
30 A/M
location in a typical commercial or
hospital environment.
meters
800 MHz to 2.5 GHz
d=[ — ]
7
P
E
1
0.23
0.74
2.3
7.4
23
EN